Rancang Bangun Peak Flow Meter dengan Output Suara dan Pemantauan Android

Patrisius Kusi Olla, Wilia Azhar

Submitted : 2021-01-11, Published : 2021-02-05.

Abstract

Peak Flow Meter (PFM) is a tool to measure the amount of air flow in the airway (PFR) and to detect asthma. The output value of PFR can be influenced by several factors, such as age, respiratory muscle strength, height and gender. In this research, airway measurements are used to measure the condition of patients suffering from asthma. The author aims to make this tool so that it can find out how to design and make a peak flow meter output sound tool, measure the peak current and can know how the MPXV7002DP sensor works in regulating output in the form of sound. The method used by the author is to design or make a tool peak flow meter output sound. This MPXV7002DP sensor works when the sensor receives air blows from the flow sensor which automatically reads the highest air pressure from the breath. The test results using the VT Mobile Medical Gas Flow Analyzer prove that the largest percentage error is 2.4%, with the blowing rate on the Peak Flow Meter is 64.0 lpm and the blowing rate on VT mobile is 62.50 lpm. Therefore, this tool can be said to be very certain to detect asthma. Then it can be concluded that the peak flow meter is feasible and meets the specified requirements.

Keywords

PFM, Asthma, Pressure, MPXV7002DP, Arduino, Bluetooth

Full Text:

PDF

References

Setiyarini, T., Abi Muhlisin, S. K. M., Kep, M., Zulaicha, E., & Kp, S. (2016). Efektivitas Pendidikan kesehatan menggunakan media leaflet dan penyuluhan individual terhadap pengetahuan pencegahan kekambuhan asma (Doctoral dissertation, Universitas Muhammadiyah Surakarta).

Lorensia, A., De Queljoe, D., & Santosa, K. A. (2017). Kelengkapan informasi mengenai cara penggunaan peak flow meter yang diberikan kepada pasien asma di apotek. Jurnal Ilmiah Manuntung, 1(2), 200-206.

Suryowinoto, A., Hamid, A., & Desmalasa, A. F. (2017). Deteksi Dini Penyakit Pernafasan Asma Dengan Peak Expiratory Flow Meter Berbasis Microcontroller. Jurnal Mikrotek, 2(4).

Anisa, A., Hamzah, T., & Mak'ruf, M. R. (2020). Peak Flow Meter with Measurement Analysis. Indonesian Journal of electronics, electromedical engineering, and medical informatics, 2(3), 107-112.

Adeniyi, B. O., & Erhabor, G. E. (2011). The peak flow meter and its use in clinical practice. Afr J Respir Med, 6(2), 5-7.

Gregg, I., & Nunn, A. J. (1973). Peak expiratory flow in normal subjects. Br Med J, 3(5874), 282-284.

Setyaningtyas, L., Herawati, I., & Fis, S. (2016). Penatalaksanaan Fisioterapi Pada Penderita Asma Bronkiale Di Rumah Sakit Khusus Paru Respira Yogyakarta (Doctoral dissertation, Universitas Muhammadiyah Surakarta).

Novarin, C., Murtaqib, M., & Widayati, N. (2015). Pengaruh Progressive Muscle Relaxation terhadap Aliran Puncak Ekspirasi Klien dengan Asma Bronkial di Poli Spesialis Paru B Rumah Sakit Paru Kabupaten Jember (The Effect of Progressive Muscle Relaxation on Peak Expiratory Flow of Clients with Bronchial As. Pustaka Kesehatan, 3(2), 311-318.

Sundaru, H. (2007). Apa yang Diketahui Tentang Asma. Departemen Ilmu Penyakit Dalam, FKUI/RSCM, 4.

ISKANDAR, S. (2011). Faktor-Faktor Risiko Yang Berpengaruh Terhadap Kejadian Asma Pada Anak Di Kota Semarang (Doctoral dissertation, Diponegoro University).

Arvin, B. K. (2000). Ilmu kesehatan anak. EGC.

Package, T. P. S. M., & Gauge, P. S. S. S. S. (2005). MPXV7002 Integrated Silicon Pressure Sensor On-Chip Signal Conditioned, Temperature Compensated and Calibrated.

SETIAWAN, E. J. (2017). Rancang bangun IoT temperature controller untuk enclosure BTS berbasis microcontroller wemos dan android (Doctoral dissertation, Universitas Mercu Buana).

Prayogo, I., Alfita, R., & Wibisono, K. A. (2017). Sistem Monitoring Denyut Jantung Dan Suhu Tubuh Sebagai Indikator Level Kesehatan Pasien Berbasis Iot (Internet Of Thing) Dengan Metode Fuzzy Logic Menggunakan Android. Jurnal Teknik Elektro dan Komputer TRIAC, 4(2), 33-39.

Swaminathan, S., Venkatesan, P., & Mukunthan, R. (1993). Peak expiratory flow rate in south Indian children. Indian pediatrics, 30(2), 207-11.

Article Metrics

Abstract view: 844 times
Download     : 463   times

Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 International License.

Refbacks

  • There are currently no refbacks.